Fluke 705 sløjfekalibrator Image #0

Fluke 705 sløjfekalibrator

  • EAN no.: 0095969062565
  • LM no.: 7898100500
  • Supplier: FLUKE DANMARK A/S
Must login for adding to cart

Description

Type no: 665091

Promotion code: 80160869054

UNSPSC (v. 19.0501): 41113630

HS Code: 90303100

For a powerful loop calibrator at a very affordable price, choose the Fluke 705 Loop Calibrator. This feature-rich solution to loop calibration, maintenance and repair offers

Includes:
AC70A Alligator Clips
C10 protective yellow holster with Flex-Stand™
Instruction sheet
TL75 safety-designed test lead set
NIST- traceable calibration report and data

Specifications
Current measurement Range 0 to 24 mA
Resolution 0.001 mA
Accuracy 0.02% Rdg + 2 LSD
Current sourcing Range 0 to 20 mA or 4 to 20 mA
Accuracy 0.025% Rdg + 2 LSD
Drive capability 1000 Ω @ 24 mA
Current simulation Range 0 to 20 mA or 4 to 20 mA
Accuracy 0.025% Rdg + 2 LSD
Ext. loop voltage requirement 12-30 V
Loop power While measuring mA, 24 V
Voltage measurement Range 0 to 28 V DC
Resolution 1 mV
Accuracy 0.025% Rdg + 2 LSD
Environmental Specifications
Operating temperature -10°C to 55°C
Storage temperature -40°C to 60°C
Humidity (without condensation) 0 to 90% (0°C to 35°C); 0 to 70% (35°C to 55°C)
Temperature coefficient -10°C to 18°C, 28°C to 55°C, ±.005% of range per°C
Safety Specifications
Safety CSA C22.2 No. 1010.1:1992
EMC EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage 30 V

Also avalibel in a EX version (fluke 707ex)

Specifications

Frequency and pulse groups encoding
FALSE
Thermal element RTD Ni1000
FALSE
Lowest resolution power sink
0.001 mA
Power sink (transmitter simulator) 0 mA ... 24 mA
TRUE
Resistor output
FALSE
Thermal element RTD Ni100
FALSE
Thermal element RTD Pt1000
FALSE
Thermal element RTD Pt100
FALSE
Calibration software
FALSE
Simultaneous measuring and encoding
TRUE

Manuals and guides

Declaration of Conformity or Declaration of Performance